Technical guideFree to readYear not stated

AFLO-TTP Portable Ultrasonic Flowmeter: User Manual

Original title AFLO-TTP Portatif Ultrasonik Debimetre: Kullanma Kılavuzu

Aktek

Aktek's AFLO-TTP clamp-on transit-time ultrasonic flowmeter manual: DN15-DN6000 pipes, ±1% accuracy (0.2 m/s and above), installation checks with signal strength 500-999 and Q value 600-900.

On this page

Overview

The manual explains use of the AFLO-TTP, a portable clamp-on transit-time ultrasonic flowmeter that measures liquid velocity in closed pipes without contact. Two transducers are clamped on the outside of the pipe; the difference in transit time between upstream and downstream directions (delta T) is proportional to fluid velocity. The mounting method depends on pipe and fluid: V method is the most common for daily measurements on 15-400 mm inner diameter pipes, Z method for above 200 mm and W method for 15-50 mm plastic pipes; N method is rare. Typical applications include cooling water, drinking water, ultra-pure liquids, chemicals, river water and plant effluent.

Technical data: linearity 0.5%, stability 0.2%, accuracy ±1% of reading (at 0.2 m/s and above), velocity ±32 m/s, pipe size DN15-DN6000, standard transducers rated to 90 °C, built-in datalogger of over 2,000 data lines, RS232 interface (75-57600 baud), over 12 hours on internal Ni-MH battery, and 4-20 mA, frequency, pulse and alarm outputs. Configuration values are retained in flash memory for over 100 years.

As measurement site, a straight pipe completely filled with the measured liquid is recommended; 10D to 30D upstream and 5D downstream straight length are needed depending on pipe configuration. A new, clean pipe should be chosen where possible; pipes with plastic liners make measurement very difficult. Transducers are fitted with acoustic gel leaving no gaps, mounted on the horizontal plane to avoid gas bubble accumulation.

Installation checks: signal strength of 500-999 is good and if below 700 the site should be changed; signal quality Q is normally 600-900; total transit time should be stable and delta time variation should not exceed 20%; the ratio of measured to calculated transit time must be 100±3, otherwise pipe parameters, transducer spacing and orientation are checked. The manual also covers menu windows, totalisers, analog and digital outputs, error codes, the communication protocol and sound velocity tables for liquids and solids.

Key points

  1. Clamp-on transit-time ultrasonic flowmeter; non-contact and unaffected by system pressure or fouling.
  2. Pipe size DN15-DN6000; velocity ±32 m/s; accuracy ±1% of reading (0.2 m/s and above); linearity 0.5%.
  3. Standard transducers are rated to 90 °C; the built-in datalogger stores over 2,000 data lines.
  4. V method for 15-400 mm, Z method above 200 mm, W method for 15-50 mm plastic pipes.
  5. Measurement site: full straight pipe; 10D-30D upstream and 5D downstream straight length.
  6. Signal strength 500-999 is good; below 700 the site should be changed.
  7. Q value should normally be 600-900.
  8. Ratio of measured to calculated transit time must be 100±3; delta time variation must not exceed 20%.
  9. Outputs: 4-20 mA, frequency, totaliser pulse, alarm, OCT and RS232; over 12 hours on internal battery.

Using it in design

  • For cooling water circuits, define measurement points beforehand for flow verification and commissioning; leave 10D-30D upstream and 5D downstream straight pipe for a clamp-on meter.
  • Check that the measurement pipe is completely full, vertical upward, inclined or horizontal, and that its liner does not block ultrasonic transmission.
  • Add signal strength, Q value, delta time and the measured-to-calculated transit time ratio (100±3) to the commissioning report.
  • Specify a high-temperature transducer in the specification for lines above 90 °C.
  • For continuous monitoring, evaluate connecting the 4-20 mA or RS232 output to the BMS or data acquisition.

Who should read it

Engineers and technicians measuring cooling water and process flow on site for commissioning and balancing.

Document contents

  1. Introduction
  2. Starting Measurement
  3. Selecting and Installing Transducer Locations
  4. Installation Check
  5. Menu Window Details
  6. How To
  7. Troubleshooting
  8. Communication Protocol
  9. Service
  10. Appendix: Sound Velocities for Liquids and Solids

Publication details

Authors
Aktek
Publisher
Aktek
Year
Year not stated
Document language
Turkish
Source type
Manufacturer
Access
Free to read
Licence
Copyright retained; citation only: the publisher’s terms apply to the original document
Original licence note

Official public manufacturer documentation inspected. No redistribution permission established. Metadata and original brief summaries only; local downloads are research evidence, not a redistribution package.

Citation

Aktek AFLO-TTP Portatif Ultrasonik Debimetre: Kullanma Kılavuzu. Aktek. https://www.aflometers.com/Upload/productDocument/aflo-ttp-kullanma-kilavuzu_2024_9_17_11_44_1_955.pdf

All content and rights belong to the respective authors and publishers. SoSo CAD provides only the citation, its own summary and a link to the source, and is not responsible for the sources' content.

Cooling capacity in SoSo CAD

Check zone heat loads, installed capacity and the capacity left with the largest unit out, as part of the design.

Zone heat loads and capacity balance in the SoSo CAD cooling workspace

Keep your reading list current

Once a month: new library sources, changes to Turkish regulations and upcoming data centre events.